2D unit |
Orthophos XG 3D ready |
---|
Programs |
|
|
[P1] - Standard panorama with orthoradial beam direction |
[P1L] [P1R] - Standard panorama half-sides left and right, e.g. for test exposures |
[P1C] - Panorama with a constant magnification of 1.25. For length measurements using a reference object
|
[P1A] - Panorama with artifact reduction. For patients with numerous metal fillings |
[P2] - Standard panorama without ascending rami |
[P10] - Pediatric panorama with beam field reduced in height and length to reduce the radiation dose
|
[P12] - Thick slice in the anterior tooth region, e.g., for extreme anomalies
|
[P20] - Panoramic radiography without ascending rami, e.g. for treating children |
Versions of the panoramic programs:
- Individual quadrants
- Artifact-reduced
- Constant magnification
- Quickshot for all panoramic programs
- Automatic adjustment of the rotation curve to the jaw width
- Automatic positioning of the head tilt with the occlusal bite block
|
[S1] Sinus program for the imaging of paranasal sinuses
|
[S2 - S4] - Sinus program
- Maxillary sinuses in one/two images
- Maxillary sinuses in one/two images (linear)
|
[MS1] Multi-slice in the posterior tooth region for determining the position of displaced molars
|
[TM1.1 / TM1.2] - Lateral temporomandibular joint with closed or open occlusion. For functional diagnostics
|
[TM2 - TM6] [TM1 - TM2] Temporomandibular joint lateral/axial
- With adjustable radiation angle
- With open and closed occlusion
- With a slice position
|
[BW1] - Bitewing exposure for the posterior tooth region. Optional panning left/right
|
[BW2] - Bitewing program for the anterior tooth region |
[BW10] Bitewing program, e.g. as a replacement for intraoral images (patients with pharyngeal reflex) |
|
Optional - [C1] Posterior-anterior exposure |
Optional - [C2] Anterior-posterior exposure |
Optional - [C3] Lateral exposure |
Optional - [C4] Carpus images |
Tube head |
|
X-ray generator |
60-90 kV; 3-16 mA |
Focal spot size acc. to IEC 60336 |
0.5 mm |
Total filtration acc. to IEC 60522
|
> 2.5 mm Al
|
2D Imaging
|
|
Sensor |
Digital CCD line sensor, repluggable |
Active sensor area |
138 x 6.48 mm |
Pixel size |
27 μm
|
Focus-sensor distance |
530 mm
|
Ceph Imaging (optional) |
|
Arm
|
Left- or right-sided |
Sensor |
Digital line sensor with CCD technology, repluggable |
Active sensor area |
230 x 6.48 mm |
Pixel size |
27 μm |
Focus-sensor distance |
1,714 mm |
General information
|
|
Country of Origin |
Made in Germany. Made by Dentsply Sirona.
|
Dimensions (D x W x H) |
No ceph: 1.411 x 1.28 x 2.25 m / 5'3'' x 4'2.4'' x 7' 4.5''
With ceph: 1.411 x 2.155 x 2.25 m / 5'3'' x 7'0.6'' x 7' 4.5'' |
Weight |
No ceph: 110 kg / 242 lbs
With ceph: 132 kg / 291 lbs |
Operating Conditions |
|
Air humidity
|
10% to 95% |
Temperature |
+10° to +40°C (50° to 104°F) |